An Introduction to the Light Microscope, Light Microscopy Techniques and Applications
Complete the form below to unlock access to ALL audio articles.
Some of the most fundamental processes in nature occur at the microscopic scale, far beyond the limits of what we can see by eye, which motivates the development of technology that allows us to see beyond this limit. As early as the 4th century AD, people had discovered the basic concept of an optical lens, and by the 13th century, they were already using glass lenses to improve their eyesight and to magnify objects such as plants and insects to better understand them.1 With time, these simple magnifying glasses developed into advanced optical systems, known as light microscopes, which allow us to see and understand the microscopic world beyond the limits of our perception. Today, light microscopy is a core technique in many areas of science and technology, including life sciences, biology, materials sciences, nanotechnology, industrial inspection, forensics and many more. In this article, we will first explore the basic working principle of light microscopy. Building on this, we will discuss some more advanced forms of light microscopy that are commonly used today and compare their strengths and weaknesses for different applications.
Contents
What is light microscopy?
Parts of a microscope and how a light microscope works
Simple and compound microscopes
Types of light microscopy
- Bright field microscopy
- Dark field microscopy
- Phase contrast microscopy
- Differential interference contrast microscopy
- Polarized light microscopy
- Fluorescence microscopy
- Immunofluorescence microscopy
- Confocal microscopy
- Two-photon microscopy
- Light sheet microscopy
- Total internal reflection fluorescence microscopy
- Expansion microscopy
Deconvolution in light microscopy
What's the difference between light microscopy vs electron microscopy?
Summary and conclusion
Light microscopy techniques comparison table
What is light microscopy?
Light microscopy is used to make small structures and samples visible by providing a magnified image of how they interact with visible light, e.g., their absorption, reflection and scattering. This is useful to understand what the sample looks like and what it is made of, but also allows us to see processes of the microscopic world, such as how substances diffuse across a cell membrane. Parts of a microscope and how a light microscope works
Fundamentally, a microscope comprises two subsystems: an illumination system to illuminate the sample and an imaging system that produces a magnified image of the light that has interacted with the sample, which can then be viewed by eye or using a camera system.Early microscopes used an illumination system comprising sunlight that was collected and reflected onto the sample by a mirror. Today, most microscopes use artificial light sources such as light bulbs, light-emitting diodes (LEDs) or lasers to make more reliable and controllable illumination systems, which can be tailored to a given application. In these systems, light from the source is typically collected using a condenser lens and then shaped and optically filtered before being focused onto the sample. Shaping the light is essential to achieve high resolution and contrast, and often includes controlling the sample area that is illuminated and the angles at which light impinges on it. Optical filtering of the illumination light, using optical filters that modify its spectrum and polarization, can be used to highlight certain features of a sample, to improve the visibility of weak signatures or to observe a sample’s fluorescence.
Figure 1: Basic compound microscope: Light from a source is focused onto the sample (object) using a mirror and condenser lens. Light from the sample is collected by an objective, forming an intermediate image which is imaged again by the eyepiece and relayed to the eye, which sees a magnified image of the sample. Credit: Technology Networks.
The imaging system collects illuminating light that has interacted with the sample and produces a magnified image that can be viewed (Figure 1). This is achieved using two main groups of optical elements:
- First, an objective lens that collects as much light from the sample as possible
- Second, an eyepiece lens which relays the collected light to the observer’s eye or a camera system.
The imaging system may also include elements such as apertures and filters that select certain portions of light from the sample, for example to see only light that has been scattered off the sample, or only light of a certain color or wavelength. As in the case of the illumination system, this type of filtering can be extremely useful to single out certain features of interest that would remain hidden when imaging all the light from the sample.
Overall, both the illumination and the imaging system play a key role in how well a light microscope performs. To get the best out of light microscopy in your application, it is essential to have a good understanding of how a basic light microscope works, and what variations exist today.
Simple and compound microscopes
Figure 3: Left: Simple microscope. Credit: Waters W./Auckland Museum. Reproduced under the Creative Commons Attribution 4.0 International (CC BY 4.0) license. Right: Compound microscope. Credit: Jacques from Cape Town, South Africa. Reproduced under the Creative Commons Attribution 2.0 Generic license. Cropped to highlight one instrument only.
In a compound microscope, the sample is illuminated from the bottom to observe transmitted light, or from the top to observe reflected light. Light from the sample is collected by an optical system consisting of two main lens groups: the objective and the eyepiece, whose individual powers multiply to enable much higher magnifications than those achieved by a simple microscope. The objective collects light from the sample and typically has a magnification of 40-100 x. Some compound microscopes feature multiple objective lenses on a rotating turret known as a “nose piece”, allowing the user to choose between different magnifications. The image from the objective is picked up by the eyepiece, which magnifies the image again and relays it to the user’s eye, with typical eyepieces having a magnification of 10 x. Therefore, the total magnification of a compound microscope, which is the product of the objective magnification and the eyepiece magnification, typically lies in the range of 400-1000 x.
For example, using green light with a wavelength of 550 nm and an objective with a typical NA of 0.7, a standard light microscope can resolve features down to a limit of 0.61 × (550 nm)/0.7 ≈ 480 nm, which is sufficient to observe cells (typically 10 µm size), but not enough to observe details of smaller organisms, e.g. viruses (typically 250-400 nm). To image smaller features, more advanced and expensive objectives with higher NA and shorter wavelengths can be used, but this may not be practical for all applications.
In standard compound microscopes (Figure 4a), the sample (often on a glass slide) is held on a stage that can be moved manually or electronically for higher precision, and the illumination system is in the lower part of the microscope, while the imaging system is above the sample. However, the microscope body can usually also be adapted to particular uses. For example, stereo microscopes (Figure 4b) feature two eyepieces at a slight angle to each other, allowing the user to see a slightly three-dimensional image. In many biology applications, an inverted microscope design (Figure 4c) is used, where both the illumination system and the imaging optics are below the sample stage to facilitate placing e.g., containers of cell cultures onto it. Finally, comparison microscopes (Figure 4d) were often used in forensics, for example to compare fingerprints or bullets by eye before the advent of digital microscopy, which allowed images to be saved and compared.
Figure 4: Compound microscope bodies. a) Standard upright microscope indicating (1) eyepiece (ocular lens), (2) objective turret, revolver, or revolving nose piece (to hold multiple objective lenses), (3) objective lenses, focus knobs (to move the stage) (4) coarse adjustment, (5) Fine adjustment, (6) Stage (to hold the specimen), (7) Light source (a light or a mirror), (8) Diaphragm and condenser, (9) Mechanical stage. Credit: GcG(jawp). b) Stereo microscope. Credit: GcG(jawp). c) Inverted microscope. Credit: Kitmondo LAB. Reproduced under the Creative Commons Attribution 2.0 Generic license. d) Comparison microscope. Credit: Tamasflex. Reproduced under the Creative Commons Attribution-Share Alike 3.0 Unported license.
Types of light microscopy
Bright field microscopy
Dark field microscopy
Dark field microscopy is a technique where only light that is scattered by the sample is collected. This is achieved by adding apertures that block the illuminating light from being imaged directly, such that only illumination light that is scattered by the sample is seen. In this way, dark field microscopy highlights small structures that scatter light (Figure 6), and can be very useful to reveal features not visible in BFM without having to modify the sample in any way. However, as the only light that is seen in the final image is light that has been scattered, dark field images may be quite dark and require high illumination powers, which may damage the sample.Figure 6: Bright field and dark field imaging. a) Polymer microstructure in bright field illumination. b) Dark field image of the same structure as in a), highlighting scattering from edges and surface contamination. c) Similar structures as in a) and b), covered by nanocrystals of 100-300 nm diameter. Only light scattered by the nanocrystals is observed, while the background light is strongly suppressed. Credit: Author.
Phase contrast microscopy
Phase contrast microscopy (PCM) is a technique that visualizes changes of optical phase caused by variations of the sample’s optical path length. This enables imaging of transparent samples that create little or no contrast in BFM, such as cells (Figure 7), for example. Because optical phase shifts are not readily observable by eye, phase-contrast microscopy requires additional optical components that convert phase shifts caused by the sample into visible changes of brightness in the final image. This requires manipulation of both the illumination system and the imaging system using apertures and filters. These shape and selectively phase shift both the light from the sample, which carries the phase information of interest, and the illumination light such that they constructively interfere on the eye or detector to create a visible image.Figure 7: Phase contrast microscopy of a human embryonic stem cell colony. Credit Sabrina Lin, Prue Talbot, Stem Cell Center University of California, Riverside.
Differential interference contrast microscopy
Differential interference contrast microscopy (DICM) enables visualization of transparent samples such as live, unstained cells by converting optical phase due to variations of the sample’s optical path length to visible contrast by interference, similarly to PCM. However, compared to PCM, DICM can achieve higher resolution images and clarity and image artefacts introduced by the optics required for PCM are reduced. In DICM, the illuminating beam is polarized by a linear polarizer and its polarization rotated such that it is split into two polarized beams with perpendicular polarization and a small (typically below 1 µm)3 separation between them. After traversing the sample, both beams are recombined such that they interfere with each other. This creates an image with contrast proportional to the difference in optical phase between the two polarized beams, hence the name “differential” interference microscopy. Images produced by DICM appear three-dimensional related to the direction of displacement between the sampling beams, which leads to the edges of the sample having bright or dark areas depending on the sign of the optical phase difference between the two (Figure 8).Figure 8: Differential interference contrast microscopy. Left: Schematic setup for DICM. Right: Live adult Caenorhabditis elegans (C. elegans) nematode imaged by DICM. Credit: Bob Goldstein, Cell Image Library. Reproduced under a Creative Commons Attribution 3.0 Unported license (CC BY 3.0).
Polarized light microscopy
In polarized light microscopy, the sample is illuminated with polarized light and the detection of the light is also sensitive to polarization. To achieve this, polarizers are used to control both the illumination light polarization and to restrict the polarization detected by the imaging system to only one specific polarization. Often, the illumination and detection polarizations are set to be perpendicular so that unwanted background illumination light that did not interact with the sample is strongly suppressed. This configuration requires a birefringent sample that introduces a rotation of the illumination light polarization angle such that it can be detected by the imaging system, for example to observe the birefringence of crystals and variations of thickness and refractive index across them (Figure 9).Figure 9: Polarization microscopy. Photomicrograph of olivine adcumulate, formed by the accumulation of crystals with different birefringence. Variations of thickness and refractive index across the sample result in different colors. Credit: R. Hill, CSIRO.
Fluorescence microscopy
Fluorescence microscopy is used to image samples that fluoresce, that is, they emit long-wavelength light when illuminated with light of a shorter wavelength. Examples include biological samples that are intrinsically fluorescent or have been labeled with a fluorescent marker, as well as single molecules and other nanoscale fluorophores. The technique employs a combination of optical filters that block short-wavelength illumination light but let longer-wavelength sample fluorescence pass, such that the final image only shows the fluorescent parts of a sample (Figure 10). This allows singling out and visualization of the distribution of fluorescent particles or cells of interest that have been stained with a dye from within a sample consisting of many other non-fluorescent particles. At the same time, fluorescence microscopy can also overcome the resolution limit of traditional optical microscopes by tagging particles that are smaller than this limit. For example viruses can be tagged with fluorescent markers to reveal their location4 or a fluorescent protein, such as green fluorescent protein,5 may be expressed in the case of biological samples. The technique has also been used to detect microplastics. Combined with various novel forms of sample illumination, this advantage of fluorescence microscopy enables “super resolution” microscopy techniques6 that break the resolution limits of traditional light microscopy. One of the main limitations of fluorescence microscopy is photobleaching, where the markers or particles cease to fluoresce because the process of absorbing the illumination light eventually alters their structure, such that they no longer emit light.Immunofluorescence microscopy
Immunofluorescence microscopy is a specific variation of fluorescence microscopy that is mainly used in microbiology to visualize the location of biomolecules within a cell. Here, antibodies tagged with fluorescent markers or that are intrinsically fluorescent bind to the biomolecules of interest, revealing their location.7Confocal microscopy
Confocal microscopy is a microscopy technique that images scattering or fluorescence from a sample point by point. Instead of illuminating and imaging the whole sample at once, an illumination spot originating from a point-like source is scanned over the sample area and only light from that point is detected with a sensitive detector, producing a 2D image. This approach allows imaging of samples with weak signals at high resolution because unwanted background signals from beyond the sampling point are efficiently suppressed. Here, the wavelength and objective used limits the size of the imaging spot in all three dimensions. This allows 2D images to be made at different depths within the sample by moving the objective to different distances from the sample. These 2D image “slices” can then be combined to create a 3D image of the sample, which is not possible with other wide field microscopy techniques discussed and also allows measurement of the sample dimensions in 3D. These advantages come at the cost of not being able to take an image in one shot, but instead having to build up the image point by point, which can be time-consuming and inhibits real-time imaging of a sample (Figure 12).Figure 12: Confocal fluorescence image of single molecule fluorescence. Small spots correspond to fluorescence from individual molecules, while larger spots correspond to clusters of molecules. The fluorescence background here is much weaker than in a simple fluorescence microscope image, as seen by the dark areas between bright spots. Credit: Author.
Two-photon microscopy
Two-photon microscopy (TPM) is a variation of fluorescence microscopy that uses two-photon absorption to excite fluorescence instead of single photon excitation. Here, fluorescence is excited by absorbing a combination of two photons of approximately half the energy required for excitation by a single photon. For example, a fluorophore that is normally excited by single blue photons could be excited by two near-infrared photons in this scheme. In TPM, the image is built up point by point just as in confocal microscopy, that is, the two-photon excitation spot is scanned across the sample and sample fluorescence detected by a sensitive detector. The large difference in excitation and fluorescence energy leads to multiple advantages compared to traditional fluorescence microscopy: first, it allows the use of longer excitation wavelengths which scatter less within the sample and thus penetrate more deeply to allow imaging of a sample below its surface and creation of 3D sample images. At the same time, photobleaching is greatly reduced as the excitation energy is much lower, which is useful for fragile samples. Fluorescence background from around the excitation spot is also greatly reduced because efficient two-photon absorption only occurs at the focus of the excitation beam, such that fluorescence from small portions of a sample can be observed (Figure 13).A disadvantage of TPM is that the probability of two-photon absorption is much lower than single-photon absorption and thus requires high-intensity illumination such as pulsed lasers to achieve a practical fluorescence signal intensity.
Light sheet microscopy
Light sheet microscopy is a form of fluorescence microscopy where the sample is illuminated by a thin “sheet” of light perpendicular to the direction of observation, such that only a thin section (typically a few micrometers) of the sample is imaged.8 By taking a sequence of images while the sample is rotated in the light sheet, a 3D image can be formed. This requires the sample to be mostly transparent, which is why this technique is often used to form 3D images of small, transparent biological structures such as cells, embryos and organisms such as zebra fish9 (Figure 14).Figure 14: Light sheet microscopy. Left: Working principle. Right: Fluorescence image of a mouse brain taken with light sheet microscopy by fluorescence imaging. Credit: Fenerliabi11, Wikimedia Commons. Reproduced under the Creative Commons Attribution-ShareAlike 4.0 International license (CC BY-SA 4.0).
Total internal reflection fluorescence microscopy
Total internal reflection fluorescence (TIRF) is a fluorescence microscopy technique that allows 2D fluorescence images to be made of an extremely thin (approximately 100 nm thick) sample slice.10 This is achieved by exciting the fluorescence of the sample by evanescent fields of the illuminating light, which occur when it undergoes total internal reflection at a boundary between two materials of different refractive index (n). Evanescent fields have the same wavelength as the illuminating light but are tightly bound to the interface. In TIRF microscopy, the excitation light typically undergoes total internal reflection at the interface between a glass slide (n = 1.52) and the aqueous medium (n = 1.35) the sample is dispersed in. The intensity of the evanescent field falls off exponentially with distance from the interface, such that only fluorophores close the interface are observed in the final image. This also leads to a strong suppression of fluorescence background from areas outside the slice, which allows weak fluorescence signals to be picked up, for example when localizing single molecules. This makes TIRF extremely useful to observe the weak signal of fluorescent proteins (Figure 15) involved in intercellular interactions, but also requires the sample to be dispersed in an aqueous medium, which may limit the types of samples that can be measured.Figure 15: TIRF image showing protein fluorescence in a cultured retinal pigment epithelial cell. Each pixel corresponds to 67 nm. Credit: Allen Liu, Sandra L. Schmid, Cell Image Library.
Expansion microscopy
The basic concept behind expansion microscopy is to increase the size of a sample that would normally require high-resolution microscopes such that it can be imaged with standard microscopy techniques, in particular fluorescence microscopy. This works well for preserved specimens such as biomolecules, cells, bacteria and slices of tissue, which can be expanded equally in all dimensions (isotropically) by up to 50 x using a chemical process11 shown in Figure 16. Expanding the samples allows isolation of individual features of interest that are normally hidden and can make them transparent allowing their interior to be imaged.Figure 16: Sample preparation for expansion microscopy. A cell is first stained and then linked to a polymer gel matrix. The cell structure itself is then dissolved (digested), allowing the stained parts to expand isotropically with the gel, allowing the stained structure to be imaged with more detail.
Deconvolution in light microscopy
Beyond adapting the optical system to a specific use case, modern light microscopy also takes advantage of digital image processing, such as image deconvolution.12 This technique allows the increase of spatial resolution as well as the localization precision of images taken with an optical microscope by compensating for blurring that is inherent to the optical system itself. This blurring can be measured in a calibration step and can then later be used to deconvolve the image, thereby reducing the blurring. By combining of high-performance optics with advanced image processing, digital microscopy13 can push the limits of resolution to see ever further into the microscopic world.Figure 17: Image deconvolution. Left: Original fluorescence image. Right: Image after deconvolution, showing increased detail. Credit: Author.
What's the difference between light microscopy vs electron microscopy?
Light microscopy typically uses wavelengths of light in the visible spectrum, which inherently limits it spatial resolution due to the Rayleigh criterion to approximately half the wavelength used (approximately 200 nm at best). However, even when using objectives with high NA and advanced image processing, this fundamental limit cannot be overcome. Instead, observing smaller structures requires the use of electromagnetic radiation of shorter wavelength. This is the underlying principle of electron microscopy, where electrons are used to illuminate the sample instead of visible light. Electrons have an associated wavelength which is much shorter than visible light, which allows magnifications of up to 10,000,000 x to be achieved, such that even single atoms can be resolved.14Figure 18: Scanning electron microscope image at 15,000 x magnification of a nanocrystal in a concentric polymer structure. Even fine details such as the pores of the substrate are resolved. Credit: Author.
Summary and conclusion
Light microscopy is a powerful tool for examining small samples across a large range of applications. By adapting the illumination and imaging technique used to the specific use case, high-resolution images can be obtained, providing insight into microscopic structures and processes in the sample. In this article, we have discussed the features, strengths and weaknesses of various light microscopy techniques, which differ in the way light impinges and is collected.Light microscopy techniques comparison table
Technique | Advantages | Limitations | Typical applications |
Bright field microscopy | Relatively simple setup with few optical elements | Low contrast, fully transparent objects cannot be imaged directly and may require staining | Imaging colored or stained samples15 and partially transparent materials16 |
Dark field microscopy | Reveals small structures and surface roughness, allows imaging of unstained samples | High illumination power required can damage the sample, only scattering image features seen | |
Phase contrast microscopy | Enables imaging of transparent samples | Complex optical setup, high illumination power required can damage the sample, generally darker images | |
Differential interference contrast microscopy | Higher resolution than PCM | Complex optical setup, high illumination power required can damage the sample, generally darker images | High resolution imaging of live, unstained cells21 and nanoparticles22 |
Polarized light microscopy | Strong background suppression from non-birefringent areas of a sample, allows measurement of sample thickness and birefringence | Requires a birefringent sample | Imaging collagen,23 revealing grain boundaries in crystals24 |
Fluorescence microscopy | Allows individual fluorophores and particular areas of interest in a sample to be singled out, can overcome the resolution limit | Requires a fluorescent sample and a sensitive detector, photobleaching can diminish signal | Imaging cell components, single molecules, proteins25 |
Immunofluorescence microscopy | Visualize specific biomolecules using antibody targeting | Extensive sample preparation, requires a fluorescent sample, photobleaching | |
Confocal microscopy | Low background signal, possible to create 3D images | Slow imaging speed, requires a complicated optical system | 3D cell imaging, imaging samples with weak fluorescence signals, surface profiling28. |
Two-photon microscopy | Deep sample penetration, low background signal, less photobleaching | Slow imaging speed, requires a complicated optical system and high-power illumination | |
Light sheet microscopy | Images only an extremely thin slice of the sample, can create 3D images by rotating the sample | Slow imaging speed, requires a complicated optical system | 3D imaging of cells and organisms8 |
Total internal reflection fluorescence microscopy | Strong background suppression, extremely fine vertical sectioning | Imaging limited to thin area of sample, requires a complicated optical system, sample needs to be in aqueous medium | |
Expansion microscopy | Increases effective resolution of standard fluorescence microscopy | Requires chemical processing of the sample, not suitable for live samples | High resolution imaging of biological samples11 |
B
References
1. Rochow TG, Tucker PA. A Brief History of Microscopy. In: Introduction to Microscopy by Means of Light, Electrons, X Rays, or Acoustics. Springer US; 1994:1-21. doi:10.1007/978-1-4899-1513-9_1
2. Smith WJ. Modern Optical Engineering: The Design of Optical Systems. McGraw-Hill; 1990. ISBN: 0070591741
3. Shribak M, Inoué S. Orientation-independent differential interference contrast microscopy. Collected Works of Shinya Inoue: Microscopes, Living Cells, and Dynamic Molecules. 2008;(Dic):953-962. doi:10.1142/9789812790866_0074
4. Gao G, Jiang YW, Sun W, Wu FG. Fluorescent quantum dots for microbial imaging. Chinese Chem Lett. 2018;29(10):1475-1485. doi:10.1016/j.cclet.2018.07.004
5. Chalfie M, Tu Y, Euskirchen G, Ward W, Prasher D. Green fluorescent protein as a marker for gene expression. Science. 1994;263(5148):802-805. doi:10.1126/science.8303295
6. Baranov M V., Olea RA, van den Bogaart G. Chasing Uptake: Super-Resolution Microscopy in Endocytosis and Phagocytosis. Trends Cell Biol. 2019;29(9):727-739. doi:10.1016/j.tcb.2019.05.006
7. Miller DM, Shakes DC. Chapter 16 Immunofluorescence Microscopy. In: Current Protocols Essential Laboratory Techniques. Vol 10.; 1995:365-394. doi:10.1016/S0091-679X(08)61396-5
8. Huisken J, Swoger J, Del Bene F, Wittbrodt J, Stelzer EHK. Optical sectioning deep inside live embryos by selective plane illumination microscopy. Science. 2004;305(5686):1007-1009. doi:10.1126/science.1100035
9. Huisken J. Slicing embryos gently with laser light sheets. BioEssays. 2012;34(5):406-411. doi:10.1002/bies.201100120
10. Fish KN. Total Internal Reflection Fluorescence (TIRF) Microscopy. Curr Protoc Cytom. 2009;50(1):273-275. doi:10.1002/0471142956.cy1218s50
11. Wassie AT, Zhao Y, Boyden ES. Expansion microscopy: principles and uses in biological research. Nat Methods. 2019;16(1):33-41. doi:10.1038/s41592-018-0219-4
12. Lam F, Cladière D, Guillaume C, Wassmann K, Bolte S. Super-resolution for everybody: An image processing workflow to obtain high-resolution images with a standard confocal microscope. Methods. 2017;115:17-27. doi: 10.1016/j.ymeth.2016.11.003
13. Hedvat C V. Digital microscopy: past, present, and future. Arch Pathol Lab Med. 2010;134(11):1666-1670. doi: 10.5858/2009-0579-RAR1.1
14. Fatermans J, den Dekker AJ, Müller-Caspary K, et al. Single Atom Detection from Low Contrast-to-Noise Ratio Electron Microscopy Images. Phys Rev Lett. 2018;121(5):56101. doi:10.1103/PhysRevLett.121.056101
15. Zhang C, Huber F, Knop M, Hamprecht FA. Yeast cell detection and segmentation in bright field microscopy. In: 2014 IEEE 11th International Symposium on Biomedical Imaging (ISBI); 2014:1267-1270. doi:10.1109/ISBI.2014.6868107
16. Nair RR, Blake P, Grigorenko AN, et al. Fine Structure Constant Defines Visual Transparency of Graphene. Science. 2008;320(5881):1308-1308. doi:10.1126/science.1156965
17. Xu D, He Y, Yeung ES. Direct Imaging of Transmembrane Dynamics of Single Nanoparticles with Darkfield Microscopy: Improved Orientation Tracking at Cell Sidewall. Anal Chem. 2014;86(7):3397-3404. doi:10.1021/ac403700u
18. Neu-Baker NM, Dozier AK, Eastlake AC, Brenner SA. Evaluation of enhanced darkfield microscopy and hyperspectral imaging for rapid screening of TiO2 and SiO2 nanoscale particles captured on filter media. Microsc Res Tech. doi:10.1002/jemt.23856
19. Li K, Miller ED, Weiss LE, Campbell PG, Kanade T. Online Tracking of Migrating and Proliferating Cells Imaged with Phase-Contrast Microscopy. In: 2006 Conference on Computer Vision and Pattern Recognition Workshop (CVPRW’06); 2006:65. doi:10.1109/CVPRW.2006.150
20. McFadzean JA, Smiles J. Studies of Litomosoides carinii by Phase-contrast microscopy: the Development of the Larvae. J Helminthol. 1956;30(1):25-32. doi:10.1017/S0022149X00032946
21. Sun W, Wang G, Fang N, Yeung ES. Wavelength-dependent differential interference contrast microscopy: selectively imaging nanoparticle probes in live cells. Anal Chem. 2009;81(22):9203-9208. doi: 10.1021/ac901623b
22. Xiao L, Ha JW, Wei L, Wang G, Fang N. Determining the full three-dimensional orientation of single anisotropic nanoparticles by differential interference contrast microscopy. Angew Chemie Int Ed. 2012;51(31):7734-7738. doi: 10.1002/anie.201202340
23. Wolman M, Kasten FH. Polarized light microscopy in the study of the molecular structure of collagen and reticulin. Histochemistry. 1986;85(1):41-49. doi:10.1007/BF00508652
24. Slámová M, Očenášek V, Vander Voort G. Polarized light microscopy: utilization in the investigation of the recrystallization of aluminum alloys. Mater Charact. 2004;52(3):165-177. doi:10.1016/j.matchar.2003.10.010
25. Lichtman JW, Conchello J-A. Fluorescence microscopy. Nat Methods. 2005;2(12):910-919. doi:10.1038/nmeth817
26. Franke W, Appelhans B, Schmid E, Freudenstein C, Osborn M, Weber K. Identification and characterization of epithelial cells in mammalian tissues by immunofluorescence microscopy using antibodies to prekeratin. Differentiation. 1979;15(1-3):7-25. doi:10.1111/j.1432-0436.1979.tb01030.x
27. Seto S, Layh-Schmitt G, Kenri T, Miyata M. Visualization of the attachment organelle and cytadherence proteins of Mycoplasma pneumoniae by immunofluorescence microscopy. J Bacteriol. 2001;183(5):1621-1630. doi:10.1128/JB.183.5.1621-1630.2001
28. Pawley J, Pawley JB. Handbook of Biological Confocal Microscopy. 2006;(August 2010). doi:10.1007/978-0-387-45524-2
29. Ellis-Davies GCR. Two-Photon Microscopy for Chemical Neuroscience. ACS Chem Neurosci. 2011;2(4):185-197. doi:10.1021/cn100111a
30. Helmchen F, Denk W. Deep tissue two-photon microscopy. Nat methods. 2005;2(12):932-940. doi:10.1038/nmeth818
31. Sako Y, Uyemura T. Total internal reflection fluorescence microscopy for single-molecule imaging in living cells. Cell Struct Funct. 2002;27(5):357-365. doi:10.1247/csf.27.357
32. Axelrod D. Total internal reflection fluorescence microscopy in cell biology. Traffic. 2001;2(11):764-774. doi:10.1034/j.1600-0854.2001.21104.x